Influence of late removal after treatment on the removal torque of microimplants

Korean Journal of Orthodontics 2022³â 52±Ç 3È£ p.201 ~ 209
±èÈ£Áø, ¹ÚÈ¿»ó,
¼Ò¼Ó »ó¼¼Á¤º¸
±èÈ£Áø ( Kim Ho-Jin ) - Kyungpook National University School of Dentistry Department of Orthodontics
¹ÚÈ¿»ó ( Park Hyo-Sang ) - Kyungpook National University School of Dentistry Department of Orthodontics

Abstract


Objective: To compare the removal torque of microimplants upon post-use removal and post-retention removal and to assess the influencing factors.

Methods: The sample group included 241 patients (age, 30.25 ¡¾ 12.2 years) with 568 microimplants. They were divided into the post-use (microimplants removed immediately after use or treatment) and post-retention (microimplants removed during the retention period) removal groups. The removal torque in both groups was assessed according to sex, age, placement site and method, and microimplant size. Pearson correlation and multiple linear regression analyses were performed for evaluating variables influencing the removal torque.

Results: The mean period of total in-bone stay of microimplants in the post-retention removal group (1,237 days) was approximately two times longer than that in the post-use removal group (656.28 days). The removal torques in the post-retention removal group (range, 4?5 N cm) were also higher than those in the post-use removal group. The mandible and pre-drilling groups demonstrated higher placement and removal torques than did the maxilla and no-drilling groups, respectively. In the no-drilling post-use removal group, the placement torque and microimplant length positively correlated with the removal torque. In the post-retention removal group, unloading in-bone stay period and microimplant diameter positively correlated with the removal torque in the no-drilling and pre-drilling methods, respectively.

Conclusions: The removal torques differed according to the orthodontic loading and removal time of microimplants. With prolonged retention of microimplants inserted using the no-drilling method, the removal torque was clinically acceptable and positively correlated with the unloading in-bone stay period.

Å°¿öµå

Microimplant; Stability; Retention; Removal torque