Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.

DLP ÇÁ¸°ÅÍ·Î Á¦ÀÛµÈ ·¹Áø Àӽüöº¹¹°ÀÇ 3Â÷¿øÀû Á¤È®µµ Æò°¡

A Study of Three-dimensional evaluation of the accuracy of resin provisional restorations fabricated with the DLP printer

´ëÇÑÄ¡°ú±â°øÇÐȸÁö 2020³â 42±Ç 1È£ p.35 ~ 41
°­¿ù, ÀÌÈñ°æ,
¼Ò¼Ó »ó¼¼Á¤º¸
°­¿ù ( Kang Wol ) - Daegu Health College Department of Dental Laboratory
ÀÌÈñ°æ ( Lee Hee-Kyung ) - Daegu Health College Department of Dental Laboratory

Abstract


Purpose: The purpose of this study was to evaluate the accuracy of the DLP 3D printer by conducting 3-dimensional assesment of resin provisional restorations.

Methods: The first premolar of the maxillary was prepared for the abutment. The abutment was scanned by using a scanner. The provisional restoration was designed by using CAD software. A total of 16 resin provisional restorations were produced using ZD200 and Veltz DLP 3D printer. Scanning was done of resin provisional restorations and 3-dimensional measurement was conducted for accuracy. The mean (SD) of RMS was reported for each group. Independent t-test was used to assess the statistical significance of the results. All analyses were done using SPSS 22.0.

Results: The mean ¡¾ SD of RMS value for the accuracy of the resin provisional restorations that was fabricated by using ZD200 and Veltz DLP 3D printer were 50.85.¡¾4.64§­ and 70.33¡¾6.31§­. Independent t-test showed significant differences between groups(p<0.001).

Conclusion: The resin provisional restorations made with DLP 3D printers showed clinically acceptable accuracy.

Å°¿öµå

Accuracy; Crowns; Printing

¿ø¹® ¹× ¸µÅ©¾Æ¿ô Á¤º¸

µîÀçÀú³Î Á¤º¸

KCI