A Study on the Loosening of the Abutment Screw of Interchangeable Implan

´ëÇÑÄ¡°úÀ̽ÄÇÐȸÁö 1998³â 17±Ç 1È£ p.12 ~ 22
¿ÀÇöÁÖ, ±ÇÇõ½Å, Á¤ÀçÇö,
¼Ò¼Ó »ó¼¼Á¤º¸
¿ÀÇöÁÖ (  ) - Á¶¼±´ëÇб³ Ä¡°ú´ëÇÐ Ä¡°úº¸Ã¶Çб³½Ç
±ÇÇõ½Å (  ) - Á¶¼±´ëÇб³ Ä¡°ú´ëÇÐ Ä¡°úº¸Ã¶Çб³½Ç
Á¤ÀçÇö (  ) - Á¶¼±´ëÇб³ Ä¡°ú´ëÇÐ Ä¡°úº¸Ã¶Çб³½Ç

Abstract


This study was aimed at evaluating the screw stability between inter changeable abutment and fixture by means of examining the loosening of abutment screw. Diameter 3.8mm thread type of pure titanium fixture in Steri-Oss system was chosen, and it had 0.75mm height of external hexed structure on its top. HL Hexed and HL conical abutment were selected in Steri-Oss, E/P and Gold UCLA abutment were selected in 3i and Estheticone and UCLA abutment were selected in Branemark system. Fixtures were mounted in the epoxy resin block and superstructures we re constructed with the above-mentioned abutments. To test screw stability under the condition of loading, implant-superstructure combinations were tightened to the torque of lONcm and the load of 6kgf and 1,000cycles was applied,
and then the torque required to loosen the abutment screw was measured with the torque gauge.

The obtained results were as follows

1.The screw stability of Steri-Oss HL Hexed abutment was 1.77 ¡¾ 0.15Ncm, and that of HL Conical abutment was 1.83 ¡¾ 0.06Ncm.
2.The screw stability of 3i E/P abutment was 2.00 ¡¾ 0.17Ncm, and that of 3i Gold UCLA abutment was 2.03¡¾0.O6Ncm.
3.The screw stability of Branemark Estheticone abutment was 0.73 ¡¾0.l5Ncm, and that of
Brahemark UCLA abutment was 1.73 ¡¾ 0.12Ncm.
4.Each screw stability was analyzed by one way ANOVA and compared by sheff test. In
result, there was significant difference between Bri nemark Estheticone abutment and
other abutment(P<0.01).