Effect of the amount of battery charge on tube voltage in different hand-held dental x-ray systems

Imaging Science in Dentistry 2012³â 42±Ç 1È£ p.1 ~ 4
±èÀº°æ,
¼Ò¼Ó »ó¼¼Á¤º¸
±èÀº°æ ( Kim Eun-Kyung ) - Dankook University College of Dentistry Department of Oral and Maxillofacial Radiology

Abstract


Purpose: Hand-held dental x-ray system is a self contained x-ray machine designed to perform intraoral radiography with one or two hands. The issue about its usage as general dental radiography is still in dispute. The aim of the present study was to assess the relationship between the amount of battery charge and the tube voltage in different hand-held dental x-ray systems.

Materials and Methods: Seven hand-held dental x-ray units were used for the study. Tube voltage was measured with Unfors ThinX RAD (Unfors Instruments AB, Billdal, Sweden) for 3 consecutive exposures at the different amount of battery charge of each unit. The average and the deviation percentage of measured kV from indicated kV of each unit were calculated.

Results: Tube voltage of only 1 unit was 70 kV (indicated by manufacturer) and those of the others were 60 kV. Tube voltage deviation percentage from the indicated kV at the fully charged battery was from 2.5% to -5.5% and from -0.8% to -10.0% at the lowest charged battery.

Conclusion: Tube voltages of all units decreased as the residual amount of the battery charge decreased. It is suggested that the performance test for hand-held x-ray system should be performed for the minimum residual charged battery as well as the full charged one. Persistent battery charging is suggested to maintain the proper tube voltage of the hand-held portable x-ray system.

Å°¿öµå

Radiography; Dental; Occupational Exposure; Radiation; Ionizing; Radiation Protection