Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density

Imaging Science in Dentistry 2020³â 50±Ç 2È£ p.153 ~ 159
Gaeta-Araujo Hugo, Nascimento Eduarda Helena Leandro, Brasil Danieli Moura, Madlum Daniela Verardi, Haiter-Neto Francisco, Oliveira-Santos Christiano,
¼Ò¼Ó »ó¼¼Á¤º¸
 ( Gaeta-Araujo Hugo ) - University of Campinas Piracicaba Dental School Department of Oral Diagnosis
 ( Nascimento Eduarda Helena Leandro ) - University of Campinas Piracicaba Dental School Department of Oral Diagnosis
 ( Brasil Danieli Moura ) - University of Campinas Piracicaba Dental School Department of Oral Diagnosis
 ( Madlum Daniela Verardi ) - University of Campinas Piracicaba Dental School Department of Oral Diagnosis
 ( Haiter-Neto Francisco ) - University of Campinas Piracicaba Dental School Department of Oral Diagnosis
 ( Oliveira-Santos Christiano ) - University of Sao Paulo School of Dentistry of Ribeirao Preto Department of Stomatology, Public Oral Health and Forensic Dentistry

Abstract


Purpose: This study was conducted evaluate the influence of reconstruction parameters of micro-computed tomography (micro-CT) images on bone mineral density (BMD) analyses.

Materials and Methods: The sample consisted of micro-CT images of the maxillae of 5 Wistar rats, acquired using a SkyScan 1174 unit (Bruker, Kontich, Belgium). Each acquisition was reconstructed following the manufacturer's recommendations (standard protocol; SP) for the application of artifact correction tools (beam hardening correction [BHC], 45%; smoothing filter, degree 2; and ring artifact correction [RAC], level 5). Additionally, images were reconstructed with 36 protocols combining different settings of artifact correction tools (P0 to P35). BMD analysis was performed for each reconstructed image. The BMD values obtained for each protocol were compared to those obtained using the SP through repeated-measures analysis of variance with the Dunnett post hoc test (¥á=0.05).

Results: The BMD values obtained from all protocols that used a BHC of 45% did not significantly differ from those obtained using the SP (P>0.05). The other protocols all yielded significantly different BMD values from the SP (P<0.05). The smoothing and RAC tools did not affect BMD values.

Conclusion: BMD values measured on micro-CT images were influenced by the BHC level. Higher levels of BHC induced higher values of BMD.

Å°¿öµå

Artifacts; Bone Density; X-Ray Microtomography