Abstract


Objective: The purpose of this study was to measure the temperature of the plugger tip of 3 cordless heat carriers set at 200¡ÆC.

Materials and Methods: Pluggers of the same taper (0.06, 0.08, 0.10) and similar tip sizes (sizes of 50 and 55) from 3 cordless heat carriers, namely SuperEndo-¥á2 (B & L Biotech), Friendo (DXM), and Dia-Pen (Diadent), were used and an electric heat carrier, System B (SybronEndo), was used as the control. The plugger tips were covered with customized copper sleeves, heated for 10 seconds, and the temperature was recorded with a computerized measurement system attached to a K-type thermometer at room temperature (n = 10). The data were analyzed with 2-way analysis of variance at a 5% level of significance.

Results: The peak temperature of the plugger tips was significantly affected by the plugger taper and by the heat carrier brand (p < 0.05). The peak temperature of the plugger tips was between 177¡ÆC and 325¡ÆC. The temperature peaked at 207¡ÆC?231¡ÆC for the 0.06 taper pluggers, 195¡ÆC?313¡ÆC for the 0.08 taper pluggers, and 177¡ÆC?325¡ÆC for the 0.10 taper pluggers. Only 5 of the 12 plugger tips showed a temperature of 200¡ÆC ¡¾ 10¡ÆC. The time required to reach the highest temperature or 200¡ÆC ¡¾ 10¡ÆC was at least 4 seconds.

Conclusion: When using cordless heat carriers, clinicians should pay attention to the temperature setting and to the activation time needed to reach the intended temperature of the pluggers.